Kalibracija radiohromatskog filma zasnovana na gradijentnim radijacionim poljima
Creator
Pecić, Stevan S., 1996-
CONOR:
130279689
Copyright date
2024
Object Links
Select license
Autorstvo-Nekomercijalno-Bez prerade 3.0 Srbija (CC BY-NC-ND 3.0)
License description
Dozvoljavate samo preuzimanje i distribuciju dela, ako/dok se pravilno naznačava ime autora, bez ikakvih promena dela i bez prava komercijalnog korišćenja dela. Ova licenca je najstroža CC licenca. Osnovni opis Licence: http://creativecommons.org/licenses/by-nc-nd/3.0/rs/deed.sr_LATN. Sadržaj ugovora u celini: http://creativecommons.org/licenses/by-nc-nd/3.0/rs/legalcode.sr-Latn
Language
Serbian
Cobiss-ID
Theses Type
Doktorska disertacija
description
Datum odbrane: 29.05.2025.
Other responsibilities
Academic Expertise
Prirodno-matematičke nauke
Academic Title
-
University
Univerzitet u Beogradu
Faculty
Fizički fakultet
Alternative title
Radiochromic film calibration based on gradient radiation fields
Publisher
[S. Pecić]
Format
100 str.
description
Fizika - Primenjena fizika / Physics - Applied Physics
Abstract (sr)
Ova disertacija istražuje razvoj alternativnog pristupa kalibraciji radiohromatskog filma, fokusirajući se na upotrebu gradijentnih radijacionih polja...
Abstract (en)
This dissertation explores the development of an alternative approach to radiochromic
film calibration, focusing on the use of gradient radiation fields. Compared to uniformly
irradiated film, the advantage of the gradient is that one exposure contains a large number
of different and previously known dose values, allowing for more efficient calibration.
The methodology developed in this work includes experimental and numerical calibration
techniques with an emphasis on the application of a physical wedge filter as the basic dose
gradient. The dynamic range coefficient is defined, the evaluation of which indicates a significant
influence of parameters such as depth and field size, while the influence of beam quality
is negligible. The measurement uncertainty of the calibration for the case of the gradient field
was examined, where the maximum deviations were defined and the total uncertainty model
obtained by uncertainty propagation was presented. Taking into account the importance of
the film scanning process, the topics of spatial resolution and the effect of lateral response
artifacts were discussed in particular.
In order to extend the calibration range, the methodology shows possibilities of extrapolation
and combination of multiple gradients. Through the formalism of the described method,
the possibilities of optimizing the positioning of the gradients for a more even distribution of
calibration points were analyzed, with a demonstration of the optimization of the procedure
through a histogram of calibration points. Optimization of the gradient placement enabled
convergence for seven exposures in the range [1, 10] Gy and eight exposures in the range
[1, 20] Gy. Based on the existing results, an algorithm for the general case of gradient field
calibration is presented.
The methodology has the potential to be implemented within standard clinical settings,
enabling routine film measurement and simplifying quality control procedures. Automation
of the protocol opens the possibility of wide application, especially for complex gradient fields
characteristic of modern therapeutic sources.
Authors Key words
Radiohromatski film, Radijaciona dozimetrija, Medicinska fizika
Authors Key words
Radiochromic film, Radiation dosimetry, Medical physics
Classification
544.2:539.16(043.3)
Type
Tekst
Abstract (sr)
Ova disertacija istražuje razvoj alternativnog pristupa kalibraciji radiohromatskog filma, fokusirajući se na upotrebu gradijentnih radijacionih polja...
“Data exchange” service offers individual users metadata transfer in several different formats. Citation formats are offered for transfers in texts as for the transfer into internet pages. Citation formats include permanent links that guarantee access to cited sources. For use are commonly structured metadata schemes : Dublin Core xml and ETUB-MS xml, local adaptation of international ETD-MS scheme intended for use in academic documents.